Download MAGNETIC BEADS

Transcript
 satisfactorily in spin columns with sample of high concentration (>250 µg/mL) over a narrow window of
sample volumes. The proprietary Zeba Micro Spin Desalting columns has exceptional protein recovery and
desalting characteristics with >95% retention of salts and other small molecules of <1,000 MW and are
recommended for processing compounds >7,000 MW.”
What is the purpose of the diafiltration spin columns in many of your kits?
These columns are used to concentrate antibody solutions. These size exclusion columns contain a
membrane of MWCO of 5,000 (this varies by the type of kit). Molecules less than this size will pass through
the membrane and larger molecules will be retained. These columns can be used to concentrate antibodies
prior to labeling or modification.
What is the purpose of the Q spin filters used in your HRP kit?
These are ion exchange columns with a positive charge on the resin surface. The pH of one of the solutions
in the kit is used to alter the charge on the conjugated antibody - above the pI of the antibody, the
antibody-HRP complex will have more negative charges, and will bind to the resin. At this pH, the HRP is
still positively charged (a higher pI) and will pass through the column, thereby separating the free HRP
from the Ab-HRP conjugate. Then, the pH is changed again, below the pI of the antibody, thereby
increasing the positive charge on the Ab-HRP conjugate, such that it passes through the column. The same
strategy is used in the PE kit.
My DMF appears to be cloudy or have a precipitate in it. Is it OK to use?
DMF is a clear, colorless liquid. The gray grainy particles (sometimes brownish-orange, depending on the
batch and supplier) are the molecular sieves that help keep it anhydrous.