Purge samples of HBr
Note: Come in early and perform this step first, the sample needs to dry again before it is picked up in
1 N HNO3 .
1. Clean the outside of the sample vials, removing any HBr drips.
2. Take samples and place in fume cupboard.
3. Pour a small amount (∼ 1 ml) of QD HNO3 into a clean vial.
4. Using the 1000µl pipette, add a few drops of QD HNO3 to each sample.
5. Cap the vials and transfer them to a drying cabinet to dry down.
Prepare TRU Spec columns
1. Clean column stand with MQ water.
2. Place waste cups underneath the column slots.
3. Take clean, dry column caps from appropriately labelled bag and place one over each column slot.
4. Pour the acid off the columns into the sink in the fume hood, with the water running. The columns
are stored in dilute HNO3 in a plastic container, which is on the hotplate in the fume hood.
5. Rinse columns in container a few times using MQ water from tank next to sink.
6. Taking the columns one by one from the container, check that the frit is in the right position (1/3
of the way down) and that the column has only a small collar/no collar and thus will sit securely
in the slotted column rack.
7. Clean the columns thoroughly with MQ water, using a squirt bottle to force water through the
frit from both top and bottom. To avoid air bubbles getting trapped in the column squirt in MQ
water from the bottom, quickly flip over and fill from top. Shake dry and repeat.
8. Fill the column with water such that no air bubbles are included.
9. Take 1000µl pipette and set it to 350µl. While there is still water in the column’s reservoir inject
the TRU Spec resin, entraining as little water as possible from the vial.
10. Once all columns have had resin added move the rack to the clean cabinet.
11. Let the resin settle onto the frit and the water drain through.
12. Once the water has drained through there should be resin sitting above the frit in a continuous
column, with no air bubbles. The resin should extend to the rim at the base of the reservoir.
13. If there are air bubbles squirt MQ water into the column from its base, re-suspending the resin
and driving out any bubbles. Place the column back in the stand to settle.
14. If there is not enough resin in the stem, re-suspend the resin by squirting MQ water up the stem
and into the reservoir above. Add more resin to the reservoir and allow to settle.
15. If there is too much resin the column, slowly add water to the reservoir to re-suspend some of the
resin. Pipette off the excess and return it to the resin vial.