Download MA100 User Manual - Motion Lab Systems, Inc.

Transcript
Linear Envelope EMG is derived from full wave rectified EMG signal. Full wave rectified
EMG represents the mathematical absolute value of the input EMG signals. The raw signal from the
EMG preamplifier has a mean value of zero as it contains both negative and positive signal
components. The process of full-wave-rectification produces an EMG signal with only positive
components by inverting all negative components of the signal such that they become positive. Thus
no information is lost and the average level of the signal will reflect the magnitude of the muscle
contractions, however it will now have a frequency bandwidth that is double that of the original
signal.
The MA-100 produces Linear Envelope EMG signals by processing the full wave rectified
EMG signal with a 40Hz low pass filter. This produces a signal that is an average of the raw EMG
activity and follows the envelope of EMG activity. Thus it will have a much lower bandwidth or
frequency content than either the raw or full wave rectified EMG signals. This signal may be further
filtered by the MA100 if necessary using the high pass filter frequencies of 5Hz, 10Hz, and 40Hz.
Select Linear Envelope EMG output by connecting the appropriate pin on the CONTROL connector
at the rear of the MA100 interface unit to ground. See page 32 for further details on selecting this
option.
2.3
Calibration and EMG output levels
Selections on the CONTROL connector are available to enable you to select different output
levels of the EMG signal from a range of ±10 volts through to as low as ±2.5 volts. The output range
of the EMG channels will be determined by the input range of the system that is to be used with your
MA100 to record or display the data. Note that changing the output level is effectively changes the
overall gain of the system.
The gain for each individual
EMG channel may be adjusted by the
small blue EMG level controls on the
back-pack. Generally you will find that
the settings of these individual EMG
level controls will be towards the upper
part of their range for normal subjects.
However they will vary depending on
the muscles under investigation and the
particular subject. Each individual
control has a gain range of x1 to x15.
This provides you with the ability to
record the maximum EMG signal level
from each muscle on a wide range of
subjects.
Fig 4 Typical EMG calibration signal (4 channels displayed)
The subject back-pack provides an integrated calibration facility. A small switch, in the center
of the back-pack (marked CAL) will inject a calibration signal (shown in Fig 4) into each EMG
channel before the gain control. This calibration signal is an 87Hz sine wave with an amplitude of
071901625
© Motion Lab Systems, Inc.
Page 12