Download MA100 User Manual - Motion Lab Systems, Inc.
Transcript
Linear Envelope EMG is derived from full wave rectified EMG signal. Full wave rectified EMG represents the mathematical absolute value of the input EMG signals. The raw signal from the EMG preamplifier has a mean value of zero as it contains both negative and positive signal components. The process of full-wave-rectification produces an EMG signal with only positive components by inverting all negative components of the signal such that they become positive. Thus no information is lost and the average level of the signal will reflect the magnitude of the muscle contractions, however it will now have a frequency bandwidth that is double that of the original signal. The MA-100 produces Linear Envelope EMG signals by processing the full wave rectified EMG signal with a 40Hz low pass filter. This produces a signal that is an average of the raw EMG activity and follows the envelope of EMG activity. Thus it will have a much lower bandwidth or frequency content than either the raw or full wave rectified EMG signals. This signal may be further filtered by the MA100 if necessary using the high pass filter frequencies of 5Hz, 10Hz, and 40Hz. Select Linear Envelope EMG output by connecting the appropriate pin on the CONTROL connector at the rear of the MA100 interface unit to ground. See page 32 for further details on selecting this option. 2.3 Calibration and EMG output levels Selections on the CONTROL connector are available to enable you to select different output levels of the EMG signal from a range of ±10 volts through to as low as ±2.5 volts. The output range of the EMG channels will be determined by the input range of the system that is to be used with your MA100 to record or display the data. Note that changing the output level is effectively changes the overall gain of the system. The gain for each individual EMG channel may be adjusted by the small blue EMG level controls on the back-pack. Generally you will find that the settings of these individual EMG level controls will be towards the upper part of their range for normal subjects. However they will vary depending on the muscles under investigation and the particular subject. Each individual control has a gain range of x1 to x15. This provides you with the ability to record the maximum EMG signal level from each muscle on a wide range of subjects. Fig 4 Typical EMG calibration signal (4 channels displayed) The subject back-pack provides an integrated calibration facility. A small switch, in the center of the back-pack (marked CAL) will inject a calibration signal (shown in Fig 4) into each EMG channel before the gain control. This calibration signal is an 87Hz sine wave with an amplitude of 071901625 © Motion Lab Systems, Inc. Page 12