Purpose:
This item sets which one of the four available edges on the cutter to use (Figure 7, see
also Figure 33). Use CUT EDGE to select a new cutting surface when the existing edge
becomes worn and will no longer make a clean cut of the tag wire. It is not necessary
to remove the cutter to change the cut edge. When changing between edges 1 or 2
and 3 or 4, the cutter must rotate 180 degrees. This rotation takes place during the next
injection cycle. When removing the cutter, the user should always note the position of
the index notch and replace it in the same orientation.
Item: TAG CREDIT
Options:
1 through 5
Purpose:
To set the number of Injector cycles allowed without a corresponding QCD cycle before
the missed tag alarm sounds.
Note: The TAG CREDIT setting does not change the QCD's detection and sorting
performance. Increasing the TAG CREDIT value simply allows the operator to
choose how many tags can be missed before the alarm sounds.
Operation:
The default TAG CREDIT is (2). Reasons to set the TAG CREDIT to a larger value are:
1) If you expect to have more than two specimens tagged before the first of those reach
the QCD.
2) If you prefer not to be alerted until more than a few tags have been missed.
Item: US-EURopean
Options:
0,000.0 or 0.000,0
Purpose:
This item sets the 1000 and decimal separator format according to the convention used
in the United States or Europe. For example, 1 million = 1,000,000 (US) = 1.000.000
(EUR). This item has no effect on the performance of the equipment.