Download ASCO-DAQ2 - NDT Ingenieros

Transcript
Combination of
channel states
Channel States
can be
Class
Comment
Combination can be:
Invalid, Valid, Warning and
Alarm
One channel, All channels
or it has no influence (n/i)
Channel numbers
1
monitoring conditions
1
2
3
4
Valid
Warn. Alarm
I
I
I
I
n/i
n/i
n/i
Invalid
V
V
V
V
n/i
n/i
n/i
Good
W
W
W
W
n/i
n/i
n/i
Warning
A
A
A
A
n/i
n/i
n/i
Alarm
I
(V, W, A)
1
All
All
All
Invalid
Possible online monitoring states are not triggered because
all channels are needed to trigger it. Instead the result “I”
leads to an a result “Invalid”
V
(V, W, A)
1
All
All
All
Good
At the end of the measurement, the valid condition is fulfilled
for all channels (all are above valid level).
V
(I, V, W, A)
1
One
All
All
Good
W
(I, V, W, A)
1
n/i
One
All
A
(I, V, W, A)
1
n/i
n/i
One
Warning At least one channel state is “warning” and not all channel
states are “alarm”. Therefore classification is Warning since
warning result have higher priority than results from the Valid
condition
Alarm
Parenthesis “()” indicate that any of the channel states can occur, no matter how often.
3.3
Complete Reference to ASCO-DAQ2 Software
3.3.1
Starting the Software
The software can be started in recording- or data replay mode. In recording mode data can be acquired. In
data replay mode previously acquired data can be replayed for analysis purposes. The first step after starting
the software is to decide whether ASCO-DAQ2 software shall be used in replay- or recording mode.
3.3.1.1 Data recording
For data recording the operator has to select a configuration file, which initializes the ASCO-DAQ2. A default
configuration is provided by Vallen Systeme.
20
file: ASCO-DAQ2_UMan_1201.doc
Copyright © 2012, Vallen Systeme GmbH