The response factor is defined as the ratio of the detector response for isobutylene to the detector response for the sample gas. Response factors for a wide range of substances have been determined experimentally. These response factors are programmed into the instrument. Note that the calibrated response curve, and all programmed response factors are relative to isobutylene. (Isobutylene has a response factor of one.) The response factor is a multiplier that compensates for the difference between the response of the sample gas and the response of isobutylene. Whenever the monitor detects a signal, it uses the response factor for that chemical to convert the signal to the correct, displayed concentration of the sample gas (if its identity is known). During calibration, this calculation is performed to define the calibrated response curve. When sampling, the isobutylene equivalent response is then multiplied by the response factor for the specific sample gas to calculate the concentration. If the response factor is known, you can use a monitor calibrated on isobutylene to calculate the actual concentration of a target gas. For example: An operator is using a monitor that has been calibrated on isobutylene. The sample gas is set to isobutylene. While using this instrument to sample for hydrogen sulfide (H2S), the display reads 100 ppm. Since the response factor for hydrogen sulfide is 3.46, the actual concentration of hydrogen sulfide is: Actual Hydrogen Sulfide concentration = 3.46 x 100 ppm = 625 ppm. Calculating a Response Factor To determine a response factor for a target chemical, perform the following simple procedure: 1. Calibrate the Sirius Detector using isobutylene as the span gas. 2. On the monitor, set the sample gas name to isobutylene. 3. Apply a known concentration of the target chemical to the monitor and note the concentration reported in the display. 4. The response factor for the target chemical relative to isobutylene: RF target gas = Actual known concentration Concentration reported by instrument 2-3